Subject: UV monitors
Peter Lundskow <plundskow [at] dnr__state__in__us> writes >Recently, we purchased a more accurate meter, the UVX Radiometer >from UVP Ultra-Violet Products (this meter comes with certification >of calibration)it takes readings in microwatts per cm squared. I >have not been able to find any information on how to convert one >measurement to the other and what safe readings would be for museums >when measured in microwatts per cm squared. Also, I would mention >our new meter has two sensor heads, one in the 250nm range and the >other in the 360nm range. I have been told by the same source that >readings with both sensors are needed to get a more complete picture >of UV exposure. You may want to consider the Hanwell Instruments ULM Universal Light Meter for taking field measurements <URL:http://www.hanwellusa.com>. It will display light levels in lux and uv levels in both microwatts/lumen and milliwatts/sq meter. This would eliminate the need to take a separate lux reading and perform the calculation to obtain the microwatt/lumen measurement. Price is $1195 USD. Following is an excerpt from the ULM manual written by Dr. Martin Hancock of Hanwell Instruments concerning design considerations and limitations of such an instrument. "The measurement of Lux and particularly UV is subject to many potential errors when performed outside laboratory conditions, a fact that is usually glossed over. This note explains some of the potential problems and gives some methods which can be used to improve the accuracy of measurements. When using equipment short of that found in a Physics laboratory, reliable measurements require a good degree of knowledge on the part of the user. Lux is a measurement of the brightness of the incident light but measured in such a way as to be comparable with the subjective opinion formed by the human eye. In practice this means choosing a detector that has a similar spectral response. Hanwell instruments use a photocell which gives an accuracy of approximately 1% fsd. It is however not a all clear that this is what is required for conservation purposes: the human eye response has no significance for objects. In the event of the spectrum of the incident light source being greatly different to the calibration source, very large errors could arise in using Lux measurements for assessing cumulative damage. Another issue is the spatial response of the sensor, in a sense that the Lux measurement should be independent of the intensity distribution of the source. This is known as cosine-correction and in practice is attempted by fitting a diffusing element in front of the photocell. In the general case of multiple light sources with different spectra, it is clear that this is not a complete solution. The problems above are even more applicable to UV measurements when there is also a requirement to provide a ratiometric measurement (which by its very nature introduces uncertainty) the end result must be interpreted with great care. To illustrate this consider the following points: * The spectral response of a UV filter is typically Gaussian-like in the spectral range 250...395 nm. This is inevitable if a 'breakthrough' form visible light (above 400 nm) is not to be included in UV measurements. * The spectral response of a typical UV detector diode is not flat in the UV region. * The pass characteristics of a suitable filter material is not flat in the UV region. These facts mean that the calibration factor varies from source to source although fortunately not by much for typical museum lighting types. The currents generated by such heavily filtered diodes at low intensities are comparable to bias and leakage currents in typical electronics, making the measurement potentially very noisy. The circuitry responds to UV power. The ratio to Lux (uW/Lumen) is then deduced afterward by hardware or software. This can lead to huge problems in practice. As an illustrative extreme example consider a Lux measurement of 1 (+/- 1 unit) and a UV power measurement of 1 (+/- 1 unit) When you take the ratio you can get any answer you like between 0 and infinity! With these problems in mind Hanwell Lux/UV instruments have been designed with the following features: * All calibration is performed in software with the original figures obtained using a 'black-body' source (Tungsten bulb). The software allows multiple calibration data sets to be entered and recalled for spectra corresponding to different colour temperatures/spectral characteristics. The relative magnitudes of these values can be found in standard texts but it is the user's responsibility to choose appropriate settings for the situation. * The circuitry uses ultra low-leakage amplifiers and capacitors to ensure a highly linear characteristic over a very wide range of UV intensities. At the lower end, the measurement (the uw/lumen ratio) is aborted when the charging rate is so low as to make significant the ratio errors described above. * As standard, a Optilas Series and SuperUV diode with a DUG11 filter material is fitted, ... but other combinations can be supplied on request to suit individual requirements." Greg Basso VP Applications Hanwell Instruments, USA 516-624-2900 *** Conservation DistList Instance 17:18 Distributed: Tuesday, August 5, 2003 Message Id: cdl-17-18-003 ***Received on Sunday, 3 August, 2003